Appendix: Using the Beckman DU-6 Spectrophotometer
June 13, 1990
Matthew S. Holt
The instructions given below are for the use of the Beckman DU-6
Spectrophotometer in quantitating and analyzing the purity of double-stranded
DNA by UV spectroscopy.
Quantitating
The Beckman DU-6 Spectrophotometer can be used to get an accurate measurement
of DNA
concentration. DNA at a concentration of 50 ug/ml has an Absorbance260 = 1.
Solving this
equation for unknown concentration, the concentration (ug/ml) = A260 x 50. To
conserve DNA
and to get accurate readings, dilute the original sample, e.g. 1:20 in 1X TE.
Hence, the final
concentration equation is:
DNA concentration (ug/ml) = A260 x Dilution Factor x 50
A 1:20 dilution is ideal for samples of concentration up to 450 ng/ul (the
concentration in ug/ml = A260 x 1000). If the DNA concentration appears to be
higher than 450 ng/ul, dilute and mix the DNA well in TE before resampling
(because the readings from too concentrated samples may not be accurate, and
are difficult to handle). Minimum diluted sample volume for quantitating is
500 ul, maximum volume is 1.5 ml.
- Turn the DU-6 Spectrophotometer on by pushing the ON/IDLE key (turn the
DU-6
off by pushing this same key). Press the UV key to turn on the UV light
source. The UV
light source needs approximately 15 minutes to warm up and will blink at the
bottom of
the screen until it has completed. Once the UV message stops blinking, the UV
light source
is ready. If the DU-6 is left at this point for over 20 minutes, a "WARNING -
LOW
ENERGY" message will appear. To correct this, turn the DU-6 off by pushing
the
ON/IDLE once to turn off, again to restart, then press the UV key and warm up
15 minutes.
- Prepare samples for quantitating. For a 1:20 dilution take 25 ul of DNA
sample and
add to 475 ul 1X TE. Samples should be vortexed hard - 15 seconds on high
speed.
You will also need 500 ul of 1X TE to use as a zero standard. Always use the
same
batch of buffer for the zero as was used to dilute the samples. If the samples
are
diluted in water, then use the same water for the zero.
- Once the UV source is ready, the DU-6 will allow parameter programming for
your
particular samples. For multi wavelength spectrophotometry press the MULTI
WAVELENGTH and the parameters will appear on the screen. Use the cursor and
the
keyboard to program the following parameters:
| MULTI WAVELENGTH | #01 |
| FUNCTION | [Abs] |
| # OF WAVELENGTHS | 3 |
| WAVELENGTH A | 230.0 |
| WAVELENGTH B | 260.0 |
| WAVELENGTH C | 280.0 |
| # OF CELLS | Variable |
Be certain to push ENTER after changing any parameter. Once all parameters
have
been entered, press the START key, and the DU-6 will start its calibrating,
which takes
approximately 2 minutes.
- After the DU-6 has finished its calibrations, the next step is to set the
zero standard.
Do this by pressing the AUTOZERO key. This tells the DU-6 to read the first
cuvette
and set that value to zero. It will only do this for the first cuvette of any
run.
- Place the zero standard in the first cuvette and the DNA samples in the
other two cuvettes.
Push RUN, and the DU-6 will print out on the top of the screen "RUNNING
SAMPLES", and on the screen the absorbance values at each wavelength for all
samples. The zero readings should be "0.000". The DU-6 will automatically
move to cuvette #2 and so on for the programmed for # of cells. After it has
read the third cuvette the DU-6 will print the message "INSERT SAMPLE - PUSH
RUN" at the top of the screen. Remove the cuvettes, aspirate thesamples, rinse
once with 1X TE, and put the next samples into the cuvettes and push "RUN".
Continue for all samples.
- When all samples have been run, push "COPY" key to get a printout of the
readings.
Rinse the cuvettes in ddH2O three times and then once in EtOH. Place open end
down on a kimwipe to dry. Press ON/IDLE key to turn off the DU-6. During
continuous quantitation, the spectrophotometer automatically prints after 14
samples have been processed to clear its screen.
Purity analysis
The DU-6 absorbance readings at various wavelengths can be used for a simple
but less indicative test of DNA purity.
- A230:A260 Ratio
Peptide bonds, amino acids, and lysis buffer salts absorb light at 210 nm.
This, however,
is an absorbance peak because many molecules absorb at 210 nm. A230 is an
absorbance
valley and should be read instead. If the sample is contaminated by peptides,
amino
acids, or lysis buffer components, an abnormally high A210 will distort A230,
giving
it a slightly higher absorbance value. The calculated ratio of A230 over A260
should
fall between 0.30 - 0.50 for pure DNA. A value higher than 0.50 indicates
possible
contamination and the DNA should be re-precipitated in EtOH and resuspended.
Once
resuspended, repeat the DU-6 spectrophotometry.
- A260:A280 Ratio
Phenol absorbs light at 270 nm. Its spectra has a slight shoulder which
appears at
approximately 272 nm. Proteins which possess tyrosine residues absorb light at
280 nm. If the sample is contaminated with any of these molecules, the
absorbance
at 280 nm will increase. The calculated ratio of A260 over A280 for pure DNA
should fall between 1.75 - 2.10. If a DNA sample has a A260:A280 ratio outside
of
these limits, reprecipitate the DNA in EtOH, resuspend, and repeat
spectrophotometry.
(If after reprecipitating this ratio is still off, this may indicate incomplete
proteinase K
digestion for genomic DNA.)
References:
ABI 340A Nucleic Acids Extractor Manual