Appendix: Operation and Maintenance of Nuaire Incubators
June 15, 1990
Rosalie Veile
Introduction:
The Nuaire incubators provide a controlled environment for optimum growth of
tissue culture cell lines. Five parameters contribute to optimum growth
conditions. These are:
- Humidity
- Precision temperature control
- Precise CO2 control
- Sterility
- Reliability
Components of the system are: The Nuaire CO2 Water-Jacketed Incubator,
the Nu-1550 Automatic Tank Switch, the Trimline Regulator, and the CO2 tanks.
1. Humidity:
Humification of the incubator is achieved through water evaporation from the
stainless steel water pan which is placed on the bottom of the shelf rack. Use
only distilled sterile mineral-free water in the stainless steel pan. The
water should be changed at least once a week. Flooding the bottm of the
incubator is not recommended, since it is difficult to change the water weekly
and almost necessitates the use of chemicals (copper sulfate; or other bacteria
growth inhibitors) which are not recommended and may damage the stainless
steel.Humidity recovery will require 15-20 minutes after door is opened for 30
seconds.
2. Precision temperature control:
37 degrees C is the set point desired, and a temperature calibration must be
performed within 1 degrees C of this temperature. To obtain the desired
temperature, place a glass mercury thermometer in a glass beaker filled with
water. Position the beaker in the middle of the incubator for several hours.
Adjust temperature reading on the display panel to coincide with temeprature of
the thermometer, and program incubator to run. Temperature of the incubator
should be checked daily.
3. Precise CO2 control:
Lymphoblastoid cell lines grow best using 5% CO2. It is necessary to determine
the proper percentage of CO2 in the chamber with the use of a fyrite. The
fyrite employs the well-known "Orsat" method of volumetric analysis involving
chemical absorption of a sample gas, as CO2. The reagent used to absorb CO2
is potassium hydroxide (dyed red). Potassium hydroxide is poisonous and care
should be taken to avoid accidental body contact or internal ingestion (on
bodyily contact, flush area with water, then wash with vinegar). To ensure
proper fyrite operation and reliable results, the fluid level should be correct
as well as the fyrite fluid strength. Refer to Bacharach fyrite gas analyzer
operation/maintenance manual for proper use and care. The autoflow incubator
uses Matheson Instruments flowmeters. The flowmeters provide accurate and low
flow measurements of CO2. To prevent dirt from the CO2 supply from interfering
with the operation of the flowmeters, a 0.3 um filter (located at the autoflow
inlet valve) should be used and changed monthly . Also, the 0.3 um air inlet
filter should be replaced monthly. The purpose of this filter is to cleanse
the room air which is drawn into the chamber. Below is a table with
approximate values to achieve the desired setting in the incubator:
Adjustments in the percentage of CO2 should be made using the CO2 flowmeter
only. The flowmeter ball will drift up and down and will not affect the CO2
percentage. It is essential that the percentage of CO2 in the chamber be
checked every two weeks or at the same time a new tank is connected to the CO2
line. If a tank is not connected correctly or there is a leak in the line it
is very important to detect the leak as soon as possible. Tubing connections
should be a tight fit and brittleness in the lines should be checked monthly.
The regulator's high pressure stage direct from the supply cylinder will have a
range from zero to 2,000 PSI. This gauge indicates actual tank pressure. The
low pressure stage should have a range from 0 to 30 PSI. This gauge will
indicate the actual CO2 pressure into the incubator system. To connect the
regulator, first open the CO2 cylinder slightly, to blow out dust or dirt that
may have collected in the valve outlet. Keep your face away from the outlet to
protect it from dust. Use some teflon tape to help seal the connection and
tighten with a wrench. Be sure the disc gasket is in place before making the
connection.
4. Sterility
The environment provided by the autoflow is not selective. As a result, any
contamination within the chamber is subjected to the same environment as the
speciments. Therefore, it is essential that all shelves and shelf brackets
should be wiped off 70% alcohol, 2% alkaline gluaraldehyde or similar
noncorrosive antimicrobial agent frequently and autoclaved monthly. Also, all
flasks placed in the incubator must be wiped with alcohol prior to being placed
into the incubator and shelves should be wiped off with alcohol every time the
cultures are fed. Mild detergent should be used to clean the outside of the
incubator. If any contamintated cultures are observed they should be isolated
from other cultures in a different incubator and fed the proper media to rid
them of contamination. The incubator in which the contaminated culture was
removed from must be sterilzed.
5. Reliability
All parts of the incubator must be reliable to maintain the optimum conditions
for tissue culture. The incubator requires periodic preventive maintenance to
insure a reliable controlled atmosphere. The CO2 supply filter and air inlet
filter should be replaced monthly. Tubing must be checked and replaced if
necessary. The chamber should be cleaned and sterilized frequently. The water
jacket requires no anti-bacterial agents because a copper tube producing copper
sulfate which eliminates bacterial growth within the water jacket is already in
place. The water jacket should be checked to insure that it is always full.
Use deionized water to fill the water jacket. Flowmeters should be checked
annually to insure CO2 percentage in the incubator (refer to preventive
maintenance, page 29 of Nuaire manual).
References:
Nuaire Incubator, Operation and Maintenance Manual, Sept. 1988.